There is a gain switch for each channel (Figure 5-6). Set the input channel gain to match the expected
voltage output of the bridge you are measuring to the input range of the A/D board as described above.
CH0
X1000
X100
X10
U
GAIN FOR CHANNELS 0 and 4
SET FOR A GAIN OF 10.
SLIDER DOWN SELECTS GAIN
ALL OTHERS TO BE OFF (UP)
O
N
O
N
CH4
Channel Gain Switches
There is a set of DIP gain switches for each input
circuit labeled GAIN (Figure 5-6). There are four,
two-position switches for each channel. The gain
switches are labeled U, X10, X100, and X1000.
Select a gain (higher than unity) by moving the
switch for that gain down. All other switches should
be left in the UP position.
A custom gain may be selected on the CIO-EXP-GP
by installing a precision resistor and setting the
switch marked “U” (User) in the down position. See
Table 5-2 below for positions and some sample gain
values.
Figure 5-6. Channel Gain Switches
Table 5-2. User Gain Resistors - Identities
RX1077RX1033
RX1066RX1022
RX1055RX1011
RX1044RX1000
Resistor PositionChannelResistor PositionChannel
10 Ohms800
17 Ohms700
40 Ohms500
161 Ohms200
364 Ohms100
776 Ohms50
Resistor ValueGain
The equation for selecting the gain resistor is:
R
USER
= (40000 / (Gain 1) ) 40
5.6 Verifying the Installation
Your channel is now configured to make thermocouple measurements. To verify the installation, use the
InstaCal program installed on your computer. This software came with your A/D board if you bought the
board from the same manufacturer as the CIO-EXP-GP. Use the CALIBRATE option to calibrate the
CJC and verify the operation of the channel. Use the TEST option to make a measurement in
engineering units.