Bode Displays
Bode Displays show vibration amplitude and phase as a function of rotating speed. You can optionally
display the Runout compensation values on the same plot by clicking the associated button (see figure).
Runout Compensation is a vector math operation, in which the referenced first order amplitude and phase
vector is subtracted from the displayed first order vector.
Note: Bode displays can only be used when one channel is selected.
1x Amplitude, selected 1xPhase selected
via button [3] via button [4]
Bode
Display
Display
Cursor
Values
Cursor
Update
(Peak
Search)
Runout
Compen-
sation
1x Phase
1x Amplitude
Bode Display, Four Plots
In this figure the plot at the lower right has focus.
In the above figure:
1xAmplitude is from button [3] in the center toolbar
1xPhase is from button [4]
To get Runout Compensation press the associated button (see figure)
7-6 Display Menu … Plots Plot Examples 937594 eZ-TOMAS