USB-2416 User's Guide Specifications
30
Table 10. E type thermocouple accuracy specifications, including CJC measurement error.
All specifications are (±).
A/D data
rate
Sensor temperature
range
Accuracy error
maximum
Accuracy error typical
Tempco
(°C/°C)
3750 S/s
-200 °C 2.413 °C 1.352 °C 0.017
0 °C 1.069 °C 0.551 °C
1000 °C 1.575 °C 1.211 °C
2000 S/s
-200 °C 2.413 °C 1.352 °C
0.017
0 °C 1.069 °C 0.551 °C
1000 °C 1.575 °C 1.211 °C
1000 S/s
-200 °C 2.413 °C 1.352 °C
0.017
0 °C 1.069 °C 0.551 °C
1000 °C 1.575 °C 1.211 °C
500 S/s
-200 °C 2.306 °C 1.244 °C
0.017
0 °C 1.017 °C 0.499 °C
1000 °C 1.539 °C 1.175 °C
100 S/s
-200 °C 2.306 °C 1.244 °C
0.017
0 °C 1.017 °C 0.499 °C
1000 °C 1.539 °C 1.175 °C
60 S/s
-200 °C 2.306 °C 1.244 °C
0.017
0 °C 1.017 °C 0.499 °C
1000 °C 1.539 °C 1.175 °C
50 S/s
-200 °C 2.306 °C 1.244 °C
0.017
0 °C 1.017 °C 0.499 °C
1000 °C 1.539 °C 1.175 °C
25 S/s
-200 °C 2.306 °C 1.244 °C
0.017
0 °C 1.017 °C 0.499 °C
1000 °C 1.539 °C 1.175 °C
10 S/s
-200 °C 2.306 °C 1.244 °C
0.017
0 °C 1.017 °C 0.499 °C
1000 °C 1.539 °C 1.175 °C
5 S/s
-200 °C 2.306 °C 1.244 °C
0.017
0 °C 1.017 °C 0.499 °C
1000 °C 1.539 °C 1.175 °C
2.5 S/s
-200 °C 2.306 °C 1.244 °C
0.017
0 °C 1.017 °C 0.499 °C
1000 °C 1.539 °C 1.175 °C
To achieve the thermocouple accuracies listed above, the USB-2416 should be warmed up for 45 minutes after
the initial power on. The accuracies listed above are only guaranteed if the USB-2416 is housed in its plastic
enclosure.