Section 4 Serial Controller
4.18
ADDRESS
The
ADDRESS
command sets the IEEE 488 bus address of the Analyzer488
when functioning as a high level controller.
The query option,
ADDRESS?,
returns the word
ADDRESS
followed by a two
digit number between 00 and 30, specifying the IEEE 488 bus address of the
Analyzer488.
TYPE High Level Controller command
SYNTAX ADDRESS {[?]|[addr]}
addr
is an IEEE 488 bus device address.
RESPONSE None
MODE High Level Controller
BUS STATES None
QUERY
ADDRESS
nn
RESPONSE
EXAMPLES
PRINT#1,"CONTROLLER ON"
select high level mode
PRINT#1,"ADDRESS 24"
set
IEEE
address to 24
PRINT#1,"ADDRESS?"
query the address
INPUT#1,A$
read
address message
PRINT A$
display
response:
ADDRESS 24