Section 4 Serial Controller
4.46
MEASURE
The
MEASURE
command is used to return the speed of previous bus events.
Responses may be returned in either a string format or a machine readable format.
If the command MEASURE
or
MEASURE0
is issued, a string in the
following format will be returned:
Time n.nnn m Seconds Count nnnnn
Bytes Rate n.nnn Bytes/Sec.
If the command
MEASURE1
is issued a machine readable string in the following
format will be returned:
nnnnnnnn mmmm,
where
nnnnnnnn
is the time in
microseconds and
mmmm
is the number of bytes transferred during that time.
Triggering, using the
SET TRIGGER
command, must be used to obtain correct
results when using the
MEASURE
command. Recording must also be on (by issuing
the
RECORD ON
command). Measuring starts when the trigger point has been found
and is stopped when the record sequence has completed. Refer to the
SET TRIGGER
command for further information on specifying trigger conditions.
In order to ensure accurate readings,
HANDSHAKE
must be set to
NONE
or
FAST
during the bus events being measured.
This command does not have a query option.
TYPE Record command
SYNTAX
MEASURE [number]
number is
0
or
1
. If not specified,
0
is assumed.
RESPONSE
Time n.nnn m Seconds Count nnnnn Bytes Rate
n.nnn Bytes/Sec.
(
MEASURE0
)
or
nnnnnnnn mmmm
(
MEASURE1
)
REMARKS To measure the response time of a particular instrument, (the time
between the device being addressed to
TALK
and when it sends out
the first byte), set the trigger conditions as follows:
MATCH
= the
talk address of the instrument,
TRIG DELAY
= 0,
POST COUNT
=