CB-7013, CB-7033 Users Manual
Calibration Sequence:
1. Connect calibration resistor to module by 4-wire R TD
connection. For CB-7033/33D, connect to channel 0.
2. Warm-Up for 30 minutes.
3. Set Type to 20- Ref. Sec .2.1.
4. Enable Calibration - Ref. Sec. 2.15.
5. Install Zero Calibration Resistor.
6. Preform Zero Calibration Command - Ref. Sec. 2.6.
7. Install Span Calibration Resistor.
8. Perform Span Calibration Command - Ref. Sec. 2.5.
9. Repeat step 4 to step 8 three times.
Note:
1. Step 4 is not needed for CB-7013/13D, version A1.x or A2.x.
2. Same for type 2A only different for set different type (step 3),
and install different Zero/Span Calibration Resistor (step 5, 7).
1.9 Configuration Tables
edoCetarduaB
300021
400042
500084
600069
edoCetarduaB
7000291
8000483
9000675
A0002511