DAQFlex Software User's Guide Hardware Reference USB-1208FS-Plus
104
Component
Supported
Property/Command
Set/Get
Supported Values
DEV
FLASHLED/{n}
Set
0 to 255
FWV
Get
MM.mm (M = major; m = minor)
ID
Set/Get
Up to 56 characters
MFGCAL
Get
yyyy-mm-dd HH:MM:SS
MFGCAL{item}
Get
YEAR as yyyy; 20xx
MONTH as mm; 01 to 12
DAY as dd; 01 to 31
HOUR as HH; 01 to 23
MINUTE as MM; 01 to 59
SECOND as SS; 01 to 59
MFGSER
Get
Up to 8 hexadecimal digits
RESET/{value}
Set
DEFAULT
DIO
Get
2 (number of digital ports)
DIO{port}
Get
8 (number of bits on the port)
DIR
Set/Get
IN, OUT (port-configurable)
LATCH
Set/Get
0 to 255
VALUE
Set/Get
0 to 255
DIO{port/bit}
LATCH
Set/Get
port number: 0, 1
bit number: 0 to 7
port value: 0 to 255
bit value: 0, 1
VALUE
Set/Get
port number: 0, 1
bit number: 0 to 7
port value: 0 to 255
bit value: 0, 1
Hardware features
8 single-ended (0 to 7), 4 differential (0 to 3) analog input channels
o Analog input ranges:
Single-ended mode: ±10V
Differential mode: ±20V, ±10V, ±5V, ±4V, ±2.5V, ±1.25V, ±1V, ±1.25V
o Sampling rate: 50 kS/s max rate
o Channel-gain queue: Up to eight unique, ascending elements; one gain element per channel.
o Resolution: 12-bit differential mode, 11-bit single-ended mode.
2 analog output channels. The analog output range is 0 V to 5 V.
16 DIO configured as two 8-bit ports. Each port is configurable for input or output.
External digital trigger
Bidirectional external clock
32-bit event counter
Vedere la pagina 103