DAQFlex Software User's Guide DAQFlex Message Reference Programming messages AOSCAN
57
HIGHCHAN
Set the last D/A channel to include in the analog output scan operation.
Message
"AOSCAN:HIGHCHAN=value"
Response
"AOSCAN:HIGHCHAN"
value
The channel number.
Example
"AOSCAN:HIGHCHAN=1"
Get the last D/A channel to include in the analog output scan operation.
Message
"?AOSCAN:HIGHCHAN"
Response
"AOSCAN:HIGHCHAN=value"
value
The channel number.
INDEX
Get the current location of the output pointer in the buffer.
Message
"?AOSCAN:INDEX"
Response
"AOSCAN:INDEX=value"
value
The current location of the pointer in the buffer.
Example
"AOSCAN:COUNT=765"
Note
This message is processed by the DAQFlex Software library, and is not sent to the
device.
LOWCHAN
Set the first D/A channel to include in the analog output scan operation.
Message
"AOSCAN:LOWCHAN=value"
Response
"AOSCAN:LOWCHAN"
value
The channel number.
Example
"AOSCAN:LOWCHAN=0'
Get the first D/A channel to include in the analog output scan operation.
Message
"?AOSCAN:LOWCHAN"
Response
"AOSCAN:LOWCHAN=value"
value
The channel number.
RANGE
Get the analog output range set for the D/A channels.
Message
"?AOSCAN:RANGE{ch}"
Response
"AOSCAN:RANGE{ch}=value"
ch
The D/A channel number.
Vedere la pagina 56