DAQFlex Software User's Guide DAQFlex Message Reference Programming messages AOSCAN
59
SCALE
Enable or disable scaling of the D/A data.
Message
"AOSCAN:SCALE=value"
Response
"AOSCAN:SCALE"
value
ENABLE, DISABLE
Example
"AOSCAN:SCALE=ENABLE"
Note
When enabled, the device accepts voltage values. When disabled, the device accepts
count values.
This message is processed by the DAQFlex Software library, and is not sent to the
device.
Get a value indicating whether scaling will be applied to the D/A data.
Message
"?AOSCAN:SCALE"
Response
"AOSCAN:SCALE=value"
value
ENABLE, DISABLE
Note
This message is processed by the DAQFlex Software library, and is not sent to the
device.
START
Start an analog output scan.
Message
"AOSCAN:START"
Response
"AOSCAN:START"
STATUS
Get the status of the AOSCAN operation.
Message
"?AOSCAN:STATUS"
Response
"AOSCAN:STATUS=value"
value
IDLE, RUNNING, or UNDERRUN
Example
"AOSCAN:STATUS=RUNNING"
STOP
Stop an analog output scan.
Message
"AOSCAN:STOP"
Response
"AOSCAN:STOP"