DataShuttle and DynaRes
11-12-01
Analog Input & Output 13-3
Select the proper thermocouple for the temperature range and accuracy required using the table below. Use
care in hostile environments especially at very high temperatures. Some thermocouples are durable only in
certain atmospheres. They may fail very quickly in another type of atmosphere. In hostile environments,
the thermocouple is usually placed in a sheath or thermowell.
Thermocouple Types
Alloy Combination
Type
+ Lead - Lead
ENickel-Chromium
(Chromel)
Copper-Nickel
(Constantan)
J Iron Copper-Nickel
(Constantan)
KNickel-Chromium
(Chromel)
Nickel-Aluminum
(Alumel)
T Copper Copper-Nickel
(Constantan)
C Tungsten 5% Rhenium Tungsten 26% Rhenium
D Tungsten 3% Rhenium Tungsten 25% Rhenium
G Tungsten Tungsten 26% Rhenium
R Platinum 13% Rhodium Platinum
S Platinum 10% Rhodium Platinum
B Platinum 30% Rhodium Platinum 6% Rhodium
N Nickel-Chromium-Silicon
(Nicrosil)
Nickel-Silicon-Magnesium
(Nisil)
Supported Product Thermocouple Ranges
Type Degrees C Degrees F
E -200 to 680 -76 to 1256
J 0 to 750 32 to 1382
K -200 to 1250 -166 to 2282
T -200 to 350 -184 to 662
C 0 to 2315 32 to 4200
D 0 to 2315 32 to 4200
G 0 to 2315 32 to 4200
R 0 to 1450 32 to 2642
S 0 to 1450 32 to 2642
B 100 to 1700 212 to 3092
N -270 to 1300 -450 to 2372