5-8 DynaRes - Technical Notes
11-13-01
DataShuttle and DynaRes
RTD Conditions Same as General Conditions. It does not include the accuracy of the RTD sensor itself.
RTD Conditions includes linearization and signal conditioning errors. Resolution assumes 16-bit resolution
is in use. RTD measurements are on the 50 mV range.
Stability
Accuracy specifications in this chapter already reflect these temperature drift errors for temperatures from
20° through 40° C.:
DynaRes 8/16 STABILITY
Range % per °C. % Over a Year
50 mV ± 0.001 ± 0.02
± 25mV ± 0.001 ± 0.02
all other voltage ranges ± 0.0025 ± 0.03
I
n many applications, stability and noise can be more important than absolute accuracy. This is
repeatability. Calculate the repeatability of a reading by adding together the stability and noise for the
range you want.
Input Impedance
The DynaRes 8/16 Ultra has an input impedance of > 20 M
.
Noise Rejection
Important: All noise measurements are in low noise mode.
CMRR The Common Mode Rejection Ratio (CMRR) is >20 dB. Conditions for CMRR: DC to 400 Hz,
common mode input ± 7.0 Volts channel to ground. This specification applies even when one or more
nonmeasured channels exceeds the operating common mode range.
DynaRes 8/16 Noise Rejection
Range Typical Internal Noise (RMS)
50 mV 0.5 µV
± 25 mV 0.5 µV
500 mV 2.5 µV
± 250 mV 2.5 µV
10 V 50 µV
± 5 V 50 µV
Common Mode Range
Operating, channel-to-ground ±7 Volts
Nonoperating 50 Volts continuous, 150 Volts for one second, nonrepetitive.
Input Protection
50 Volts continuous. 150 Volts for one second, nonrepetitive.