
48
CB-7021, CB-7022, CB-7024 User’s Manual
2.17 $AA1N
Description:
CB-7022: Perform 20 mA Calibration for Channel N
CB-7024: Perform 20 mA/+10V Calibration for Channel N
Note: The command is for CB-7022 and CB-7024 only.
Syntax: $AA1N[CHK](cr)
$ Delimiter character
AA Address of setting module (00 to FF)
1 Command for perform 20 mA (or +10V) calibration
N Channel to calibrate (0 to 1 for CB-7022, 0 to 3 for
CB-7024)
Response: Valid Command: !AA[CHK](cr)
Invalid Command: ?AA[CHK](cr)
Syntax error or communication error may get no
response.
! Delimiter for valid command
? Delimiter for invalid command
AA Address of response module (00 to FF)
Example:
Command: $0112 Receive: !01
Perform address 01 channel 1 calibration (20 mA for
CB-7022, 20 mA or 10.0V for CB-7024), returns successful.
Related Command:
Sec. 2.16 $AA0N, Sec. 2.18 $AA3NVV
Related Topics:
Sec. 1.9 Calibration
Commenti su questo manuale